iBox® Explorer Imaging Microscope Instruction Guide __________________________________________________________________
iBox Explorer2 10 Dichroic Splitter Cube Installation The dichroic splitter cube is used to direct coaxial lighting down through the magnificati
iBox Explorer2 11 Filters are installed in their black casing at the UVP factory. To insert the filter/holder into the filter wheel, hold the
iBox Explorer2 12 Platform Stage/Warming Plate The platform stage contains a built-in retractable warming plate. The warming
iBox Explorer2 13 Additional iBox Explorer2 Connections Plug the USB cable into the back of the darkroom cabinet and the other end into the com
iBox Explorer2 14 Operation Preparing the System Turn on the darkroom and the BioLite Xe prior to opening the software to enable the hardware
iBox Explorer2 15 Preparing to Capture Images Platform Stage/Warming Plate The platform stage/warming plate creates a constant temperature of 38ºC
iBox Explorer2 16 Selecting Emission Filters Filter selection is controlled via the Microscope menu. Select the appropriate emission filter from t
iBox Explorer2 17 Fine Focus While twisting the joystick handle will cause quick vertical movements of the sample platform, the “+” and “–“but
iBox Explorer2 18 Capturing Images Prior to image capture, use the Preview button to adjust image focus, exposure, location, etc. For e
iBox Explorer2 19 Service Procedures Care and Cleaning CAUTION: Warming plate surface can be hot. Let the plate cool prior to cleaning the unit. T
iBox Explorer2 2 Table of Contents Table of Contents ...
iBox Explorer2 20 Replacement Parts and Accessories Replacement parts and accessories part numbers are shown below. To order accessories or replac
iBox Explorer2 21 Technical Support UVP offers expert technical support on all UVP products. If there are any questions about product use, operat
iBox Explorer2 3 Introduction The iBox Explorer2 Imaging Microscope is designed to automate research with one-touch preset or user-defined PC cont
iBox Explorer2 4 Safety Information Do not use this equipment in a manner other than as stated in these instructions. This equipment is designed
iBox Explorer2 5 Components The iBox Explorer2 Imaging Microscope is comprised of the following components: Darkroom Cabinet Joystick CCD
iBox Explorer2 6 Specifications Power Requirements: 100-230V; 6.3 Amps at 120 Volts Mains supply voltage fluctuations are not to exceed 10
iBox Explorer2 7 Installation Installing the Software Insert VisionWorksLS CD (not network CD) into the computer. Click on the Install butto
iBox Explorer2 8 If the computer is not connected to the Internet, click Offline activation to register the software. This allows the user to o
iBox Explorer2 9 Installing the Hardware Darkroom Setup When unpacking and setting up the darkroom, two people are required to move the darkroo