Uvp Crosslinkers Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Ekwipunek Uvp Crosslinkers. UVP Crosslinkers User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 16
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
Ultraviolet Crosslinkers
Operating Instructions
and Service Manual
IMPORTANT: Please read these instructions before operating your
UVP UV Crosslinker to familiarize yourself with its operation.
UVP, LLC
2066 W. 11th Street
Upland, CA 91786
Phone: (800) 452-6788
Fax: (909) 946-3597
Web Site: www.uvp.com
81-0112-01 Rev G
Ultra-Violet Products Ltd.
Unit 1, Trinity Hall Farm Estate,
Nuffield Road, Cambridge CB4 1TG UK
Phone: +44(0)1223-420022
Fax: +44(0)1223-420561
Przeglądanie stron 0
1 2 3 4 5 6 ... 15 16

Podsumowanie treści

Strona 1 - Ultraviolet Crosslinkers

Ultraviolet Crosslinkers Operating Instructions and Service Manual IMPORTANT: Please read these instructions before operating your UVP UV

Strona 2

UV CROSSLINKERS UVP 10 14. Set the radiometer scale to 20mW/cm2. 15. Connect the radiometer to PEN 2 of the chart recorder. 16. Select 1V sc

Strona 3 - IMPORTANT SAFETY INFORMATION

UV CROSSLINKERS UVP 11 GENERAL SERVICING PROCEDURES A. Operations without Removing the Top Cover Changing the tubes It is recommended to ch

Strona 4 - OPERATION

UV CROSSLINKERS UVP 12 Note: If more than one tube holder is to be replaced, the wiring schematic diagram may be of assistance. Replacement of st

Strona 5

UV CROSSLINKERS UVP 13 3. Release main connector by holding the nuts with a spanner and then on the outside back of the unit, release the two scr

Strona 6

UV CROSSLINKERS UVP 14 C. Operations with the Internal Chassis Assembly Removed Removal of the assembly 1. Release the wire to the micro swi

Strona 7 - APPLICATIONS

UV CROSSLINKERS UVP 15 Changing plastic feet The feet are fitted to the bottom of the cabinet. They are removed by unscrewing the screw mounted

Strona 8 - CHANGING THE UV WAVELENGTH

UV CROSSLINKERS UVP 16 TECHNICAL SUPPORT UVP offers technical support on all of its products. If you have any questions about the product’s use,

Strona 9

UV CROSSLINKERS UVP 2 CONTENTS Introduction 3 Important Safety Information 3 Description and Specifications 3 Operation 4 A

Strona 10 - WITH REFLECTOR REMOVED

UV CROSSLINKERS UVP 3 INTRODUCTION UVP Crosslinkers offer researchers an instrument to quickly, safely and efficiently expose samples to a contro

Strona 11 - GENERAL SERVICING PROCEDURES

UV CROSSLINKERS UVP 4 • Internal safety interlock • Large LCD readout • Tactile membrane switch keypad • UV blocking viewing window • Large i

Strona 12 - UV CROSSLINKERS UVP

UV CROSSLINKERS UVP 5 The UV Crosslinker can be operated on the following settings: A. Preset UV Energy Exposure Setting: Push the PRESET a

Strona 13

UV CROSSLINKERS UVP 6 C. User-Defined UV Energy Exposure Setting: Sometimes it may be necessary for you to set your own standards for exposure

Strona 14

UV CROSSLINKERS UVP 7 APPLICATIONS The Crosslinker is a multi-purpose ultraviolet exposure instrument for use in the laboratory. A wide variety o

Strona 15 - SCHEMATIC DIAGRAM

UV CROSSLINKERS UVP 8 CHANGING THE UV WAVELENGTH A UV Crosslinker is purchased with shortwave, longwave or midrange tubes. However, if user requ

Strona 16 - WARRANTY

UV CROSSLINKERS UVP 9 CROSSLINKER CALIBRATION PROCEDURE Equipment necessary: • Chart recorder • Two probes with banana plug adapters • One ra